A-100nv, Page 13 – Xylem A-100NV DIGITAL PERISTALTIC PUMPS User Manual
Page 13

PROGRAM MODE 3
minimum input value
RUN
PROGRAM
STAND-BY
PRIME
MINIMUM
MAXIMUM
3
MODE
00.0
VDC
4 Enter the motor speed at the minimum VDC input
signal value. Press the FIELD button to select the
digit to program. The digit will blink when selected.
4 Press the DIGIT button to change the selected digit.
4 Repeat until all digits are programmed.
4 Press the mode button. The % SPEED icon will stop
blinking and the VDC icon will blink indicating the
minimum VDC value is ready to be programmed. The
currently programmed minimum value is shown on
the 3-DIGIT LCD.
4 Enter the minimum VDC input signal value. Note:
this value must be less than the maximum VDC input
signal value. Press the FIELD button to select the digit
to program. The digit will blink when selected.
4 Press the DIGIT button to change the selected digit.
4 Repeat until all digits are programmed.
4 Press the mode button. The VDC icon will stop
blinking and the % SPEED icon will blink. The
ARROW next to the word MAXIMUM will blink
indicating the maximum value is ready to be pro-
grammed. The currently programmed maximum
motor speed value is shown on the 3-DIGIT LCD.
4 Enter the motor speed at the maximum VDC input
signal value. Press the FIELD button to select the
digit to program. The digit will blink when selected.
4 Press the DIGIT button to change the selected digit.
4 Repeat until all digits are programmed.
4 Press the mode button. The % SPEED icon will stop
blinking and the VDC icon will blink indicating the
maximum VDC value is ready to be programmed. The
currently programmed maximum value is shown on
the 3-DIGIT LCD.
4 Enter the maximum VDC input signal value. Note:
this value must be greater than the minimum VDC
input signal value. Press the FIELD button to select
the digit to program. The digit will blink when
selected.
4 Press the DIGIT button to change the selected digit.
4 Repeat until all digits are programmed.
4 Press the mode button. Programming is complete.
4 To exit the programming mode, press the
RUN/STANDBY button and the MODE button at the
same time. The PROGRAM arrow will disappear.
PROGRAM MODE 3
% speed at the maximum input
RUN
PROGRAM
STAND-BY
PRIME
MINIMUM
MAXIMUM
% SPEED
3
MODE
100
PROGRAM MODE 3
maximum input value
RUN
PROGRAM
STAND-BY
PRIME
MINIMUM
MAXIMUM
3
MODE
10.0
VDC
A-100NV
Page 13