Cond 3110

16

ba75792e02

11/2008

4.2

Measuring

Preparatory activities

Perform the following preparatory activities when you want to measure:

Stability control

AutoRead

During the measuring procedure, the stability control function is
automatically activated. The stability control function (AR) checks the
stability of the measured conductivity signal and the stability of the
measured temperature signal. The stability has a considerable effect
on the reproducibility of the measured value.

For identical measurement conditions, the following applies:

Temperature sensor

The temperature measurement is absolutely essential for a
reproducible conductivity measurement. If a temperature sensor is
integrated in the sensor, it is indicated on the display by TP.

Note
The conductivity meter automatically recognizes the type of the
temperature sensor used. Therefore, you can connect measuring cells
with an NTC30 or Pt1000.

1

Connect a measuring cell to the meter.

2

Adjust the temperature of the test solutions and measure the
current temperature if the measurement is made without a
temperature sensor.

3

Calibrate or check the meter with the measuring cell.

4

Select the measured parameter with .

Measured
parameter

Reproducibility

Response time

Conductivity

better than 0.5% of measured
value

> 10 seconds

Temperature

< 0.3 °C of temperature value

> 15 seconds