5-12 SDI-12 Command and Response Protocol

H-3551

5.14 Extended “XTEST” Command
This command is used for installation and testing and requires the use of a H-4191 Sidekick
interface and a PC. This command causes the H-3551 to display a listing of the H-3551's current
settings, followed by a repeating printout of real-time measurement data. This is not compliant
with the SDI-12 specification and is not used with data loggers.

An example of an “XTEST” printout is shown below:

H-3551 Settings:
Firmware Checksum = PASS
Bubbler Mode = Remote Host
Bubble Rate (bub/s) = 60
Hysteresis (psi) = 0.50
Purge Pressure (psi) = 40.00
Purge Threshold (psi)= 20.00
Purge Time (sec) = 45
AZ Interval (min) = 30

Tank_PSI=xx.xxx, Tank_TEMP=xx.xx, Tank_Vp=xxxxx, Tank_Vt=xxxxx
Line_PSI=xx.xxx, Line_TEMP=xx.xx, Line_Vp=xxxxx, Line_Vt=xxxxx

Tank_PSI=xx.xxx, Tank_TEMP=xx.xx, Tank_Vp=xxxxx, Tank_Vt=xxxxx
Line_PSI=xx.xxx, Line_TEMP=xx.xx, Line_Vp=xxxxx, Line_Vt=xxxxx

Tank_PSI=xx.xxx, Tank_TEMP=xx.xx, Tank_Vp=xxxxx, Tank_Vt=xxxxx
Line_PSI=xx.xxx, Line_TEMP=xx.xx, Line_Vp=xxxxx, Line_Vt=xxxxx

etc.