H-3551

SDI-12 Command and Response Protocol 5-11

"aD0!"

"a+30.0

BubbleRate is 30.0 bubbles/min

5.10 Extended Test Pump Motor Command
This command is used during installation and testing to activate the pump motor for 5.0 seconds.

Example of a H-3551 Extended "Test Pump Motor" command:

Command

Response

Description

"aXTPM!"

"0050"

Activate the pump motor for 5 seconds

5.11 Extended Test Purge Valve Command
This command is used during installation and testing to activate the purge valve solenoid for 5.0
seconds.

Example of a H-3551 Extended "Test Purge Valve" command:

Command

Response

Description

"aXTPV!"

"0050"

Activate the purge valve for 5 seconds

5.12 Extended Test Auto-Zero Valve Command
This command is used during installation and testing to activate the auto-zero valve solenoid for
5.0 seconds.

Example of a H-3551 Extended "Test Auto-Zero Valve" command:

Command

Response

Description

"aXTAZ!"

"0050"

Activate the auto-zero valve for 5 seconds

5.13 Extended Test COP (Computer Operating Properly) Timer Command
This command causes the COP timer to expire which in turn should initiate a system reset.

Example of a H-3551 Extended "Test COP Timer" command:

Command

Response

Description

"aXCOPT!"

"0050"

Test the COP timer