H-3521 – Xylem H-3521 User Manual
Page 45

H-3521
SDI-12 Command and Response Protocol 5-15
5.14 Extended Read Measure Rate and Write Measure Rate
When the H-3521 is put in power mode 3 it makes continuous measurements based on a measure
rate. This command is for reading/writing that measure rate value. The measure rate range is 0-
255 minutes. When the user sets the measure rate to 0 then the H-3521 makes continuous
measurements as fast as the H-3521 will measure, which is about 6 seconds by default. The
factory default is 1 minute measure rate.
Command
Response
Description
"aXRMR!"
“a0011
Read Measure Rate
"aXWMRnnn!"
“a0021
Write Measure Rate
Where:
a
is the sensor address ("0-9", "A-Z", "a-z", "*", "?").
XRMR
are upper case characters.
XWMR
are upper case characters.
nnn
is the rate in minutes that the sensor will update all outputs.
Example of a H-3521 Extended "Read Measure Rate" command:
Command
Response
Time
Values
Description
"aXRMR!"
"a0011
1 sec
1
Read Measure Rate
Command
Response
Description
"aD0!"
"a+1
Measure Rate = 1 minute
Example of a H-3521 Extended "Write Measure Rate" command:
Command
Response
Time
Values
Description
"aXWMR2!"
"a0021
2 sec
1
Write Measure Rate
Command
Response
Description
"aD0!"
"a+2
Measure Rate = 2 minutes