4-4 Simple RS-232 Menu Interface

H-3521

screen will refresh and show the new offset. You must make a new measurement “M” to
see the results of the new offset.

4.3.7 U - Slope
This option is for displaying the current stage slope and also for editing the stage slope.
The following prompt at the bottom of the menu.

Enter Option >Enter Slope Value[ ]

Enter the new stage slope and press the Enter key to input the value. The screen will
refresh and show the new slope. You must make a new measurement “M” to see the
results of the new stage slope.

Note: When the stage slope is set to 2.3067 then the label (PSI to Feet) is appended to the
slope value and when the slope is set to 0.70308 then the label (PSI to Meters) appears.

4.3.8 C - Averaged Measurements
This option is for displaying the current Averaged Measurements/Meancount value and
for editing the value. The option also displays the estimated time that the measurement
will take due to the amount of averaging. The following prompt at the bottom of the
menu.

Enter Option >Averaged Measurements [ ]

Enter the new averaged measurements/meancount and press the ENTER key. Refer to
Chapter 5 for more details on the meancount value.

4.3.9 M - Measure
Pressing the M key causes the H-3521 to make a new measurement and refresh the screen
with the new measurement data. The user will see the following message after pressing
the M key.

Enter Option >Measuring...

4.3.10 X - Exit
Pressing the X key causes the H-3521 to exit the Simple Menu mode. The user will see
the following message after pressing the X key.

Enter Option >

H-3521 Fluid Off