YSI Level Scout User Manual
Page 53

YSI Incorporated
Level Scout 51
offset)
“3” is number values in “D0” resp:
asoc
“s” is signed user slope
“o” is signed user offset (psi)
“c” is signed field cal. Offset
M4
Start non concurrent 
standards lab scale 
factor measurement 
(standards lab 
calibration slope and 
calibration offset) 
aM4!
:
aD0!
a0012
:
a (service request response)
“3” is number values in “D0” resp:
aso
“s” is signed cal. Slope
“o” is signed cal. offset
M5
Start non concurrent 
high accuracy PCB 
temperature 
aM5!
:
aD0!
attt2
:
a (Service Request response)
“2” is number values in “D0” resp:
atu
“t” is signed temperature, “u” is = units: 
0-Celsius, 1=Fahrenheit. Use “XUT” 
command to set units is si 
M6
Start non concurrent 
battery voltage 
measurement 
aM6!
:
aD0!
a0021:
:
a (Service request response)
“01” is number values in “D0” resp:
av
“v” is the battery voltage
M7
Start non concurrent 
pressure and 
temperature 
measurement 
aM7!
:
aD0!
attt4
4 is number values in “D0” resp:
aputu
p is ± pressure value, first u is = units 
index. Units set by “XUP” command. t 
is ± temperature value, second u is +0 = 
Celsius or +1 = Fahrenheit. Use “XUT” 
command to set units. 
