Commissioning
TetraCon
®
700 IQ (SW)
3 - 4
ba76018e01
01/2012
Carrying out settings
Using
s
, switch from the measured value display to the main menu of
the settings. Then navigate to the setting menu (setting table) of the
sensor. The exact procedure is given in the relevant IQ S
ENSOR
N
ET
system operating manual.
Temp. compensation
linear
0.5 ... 3.0 %/K
Factor for linear temperature
compensation. This menu item only
appears if linear temperature
compensation is selected.
Reference
temperature
Tref20
(20 °C)
Tref25
(25 °C)
The reference temperature is the basis for
calculating the temperature compensation.
TDS factor
0,40 .. 1,00
Factor for automatic calculation of the total
dissolved solids by the sensor.
Cell constant
0.826 ... 1.008 cm
-1
Here you can set the cell constant if this is
necessary for a special application, e.g.
when using a flow-thru vessel.
Temperature
adjustment
-1.5 ... +1.5 K
Here you can balance the temperature
sensor in the sensor against a reference
temperature measurement.
Save and quit
The system confirms the saving of the
settings and the display switches to the
next higher level.
Quit
The display switches to the next higher
level without saving the new settings.