TetraCon
®
700 IQ (SW)
Commissioning
3 - 3
ba76018e01
01/2012
3.4
TetraCon
®
700 IQ (SW) setting table
Menu item
Selection/values
Explanations
Measuring mode
Conductivity /cm
Salinity
TDS
Conductivity /m
Measured parameter in the measured
value display
(TDS = total dissolved solids)
Measuring ranges
with
Measuring mode,
Conductivity /cm
AutoRange
0 .. 20.00
m
S/cm
0 .. 200.0
m
S/cm
0 .. 2000
m
S/cm
0 .. 20.00 mS/cm
0 .. 200.0 mS/cm
0 .. 500.0 mS/cm
These measuring ranges are available for
selection. If the AutoRange menu item is
selected, the measurement range selection
and switchover is made automatically.
Measuring ranges
with
Measuring mode,
Conductivity /m
AutoRange
0 .. 2.000 mS/m
0 .. 20.00 mS/m
0 .. 200.0 mS/m
0 .. 2000 mS/m
0 .. 20.00 S/m
0 .. 50.00 S/m
Measuring range in
Measuring mode
Salinity
0 .. 70
The measuring range is permanently set.
Measuring range in
Measuring mode
TDS
0 .. 2000 mg/l
The measuring range is permanently set.
Temperature mode
°C
°F
Unit of the measured temperature value
(Celsius, Fahrenheit).
Temp. compensation
none
nonlinear
linear with setting
For natural waters (groundwater,
surface water, drinking water),
salinity (seawater) according to IOT
Other aqueous measuring solutions