2 info messages, Info messages -2 – YSI IQ S ENSOR N ET TriOxmatic 701 IQ User Manual
Page 54
Indexes
TriOxmatic
®
701 IQ
9 - 2
ba76047e01
01/2012
9.1.2
Info messages
EI1332
Operational voltage too low
* Check installation and cable lengths, Follow installation instructions
* Power unit(s) overloaded, add power unit(s)
* Check terminal and module connections
* Defective components, replace components
EI2332
Operational voltage too low, no operation possible
* Check installation and cable lengths, Follow installation instructions
* Power unit(s) overloaded, add power unit(s)
* Check terminal and module connections
* Defective components,
replace components
ES1332
Component hardware defective
* Contact service
ESB332
SensReg: Electrolyte supply is depleted
* Change electrolyte solution and membrane head immediately (see
operating manual)
ESC332
SensLeck: Membrane head damaged
* Change electrolyte solution and membrane head immediately (see
operating manual)
Message code
Message text
Message code
Message text
IC1332
Sensor has been successfully calibrated
* For calibration data, see calibration history
IC4332
Last valid calibration has been activated. Make sure the sensor
operates correctly.
II1332
Language not available,
Default language German
* Contact service