9 indexes, 1 explanation of the messages, 1 error messages – YSI IQ S ENSOR N ET TriOxmatic 701 IQ User Manual
Page 53: Indexes -1, Explanation of the messages -1 9.1.1, Error messages -1, 9indexes

TriOxmatic
®
701 IQ
Indexes
9 - 1
ba76047e01
01/2012
9
Indexes
9.1
Explanation of the messages
This chapter contains a list of all the message codes and related
message texts that can occur in the log book of the IQ S
ENSOR
N
ET
system for the TriOxmatic
®
701 IQ sensor.
Note
Information on the contents and structure of the log book, and how to
call it up, is given in the L
OG
BOOK
chapter of the IQ S
ENSOR
N
ET
system operating manual.
Note
All message codes of the TriOxmatic
®
701 IQ end up with "332".
9.1.1
Error messages
Message code
Message text
EA1332
Meas. range exceeded or undercut
* Check process
* Select other meas. range
EA2332
Sensor temperature too high!
* Check process and application
EA3332
Sensor temperature too low!
* Check process and application
EC4332
Sensor could not be calibrated,
sensor blocked for measurement
Cause: instable signal
* Check polarization time and temperature adjustment
* Check calibration conditions
* Read calibration history
* Repeat calibration, if unsuccessful: service the sensor (see operating
manual)
EC5332
Sensor could not be calibrated,
sensor blocked for measurement
Cause: slope error
* Check polarization time and temperature adjustment
* Check calibration conditions
* Read calibration history
* Repeat calibration, if unsuccessful: service the sensor (see operating
manual)