Commissioning
3 - 1
ba76047e01
01/2012
3
Commissioning
3.1
Scope of delivery
TriOxmatic
®
701 IQ
ZBK 25i accessory case
The sensor is filled with electrolyte and provided with a protective
hood
and protective caps.
Operating manual.
3.2
Installation
Connection cable
The SACIQ sensor connection cable is required to connect the sensor.
Information on this and other IQ S
ENSOR
N
ET
accessories is given in
the YSI catalog and on the Internet.
Note
How to connect the SACIQ sensor connection cable to the terminal
strip of an MIQ module is described in chapter 3 I
NSTALLATION
of the
IQ S
ENSOR
N
ET
system operating manual.
Are the plug
connections dry?
Before connecting the sensor and sensor connection cable, please
make sure that the plug connections are dry. If moisture gets into the
plug connections, first dry the plug connections (dab them dry or blow
them dry using compressed air).
Note
Do not suspend the sensor on the sensor connection cable. Use a
sensor holder or an armature. Information on this and other
IQ S
ENSOR
N
ET
accessories is given in the YSI catalog and on the
Internet.
Minimum approach flow
The minimum required approach flow at the sensor must be present
during measurement (see chapter 7 T
). The minimum
approach flow can, e.g. be achieved by:
the flow rate of the water
turbulence in the activated sludge tanks in the vicinity of the sensor.