2 application characteristics, Application characteristics -2 – YSI IQ S ENSOR N ET TriOxmatic 701 IQ User Manual
Page 48
Technical data
TriOxmatic
®
701 IQ
7 - 2
ba76047e01
01/2012
Air pressure
compensation
7.2
Application characteristics
Allowed temperature
range
Allowed pH range of the
test sample
4 ... 12
Pressure resistance
Sensor with connected SACIQ sensor connection cable:
The sensor meets all requirements according to article 3(3) of 97/23/
EC ("pressure equipment directive").
Type of protection
Sensor with connected SACIQ sensor connection cable:
IP 68, 10 bar (10
6
Pa)
Immersion depth
min. 10 cm; max. 100 m depth
Operating position
Any, if the minimum approach flow is guaranteed
Approach flow
≥
23 cm/s (at 1 % measurement accuracy)
Fields of application
Water and wastewater monitoring
Measurement of the residual oxygen in the denitrification process
Oxygen entry measurements
Automatic
With IQ S
ENSOR
N
ET
systems
with air pressure measurement
(option)
Manual
By entering the air pressure or
location altitude
Measuring medium
0 °C ... + 60 °C (32 ... 140 °F)
Storage/transport
- 5 °C ... + 65 °C (23 ... 149 °F)
Max. allowed overpressure
10
6
Pa (10 bar)