4 cleaning the electrodes, Cleaning the electrodes -10, Es (section 5.4 – YSI IQ S ENSOR N ET TriOxmatic 701 IQ User Manual
Page 32
Maintenance, cleaning, disposal, and replacement
TriOxmatic
®
701 IQ
5 - 10
ba76047e01
01/2012
5.4
Cleaning the electrodes
The counter electrode and reference electrode always exhibit a distinct
coloration. This is necessary for the operation of the sensor. This is not
contamination. Cleaning is only required in cases of slopes that are too
small or too large (the sensor cannot then be calibrated) that cannot be
resolved by changing the membrane head and the electrolyte solution.
Caution
To clean the electrodes, always unscrew the sensor from the
SACIQ sensor connection cable. Otherwise, undesirable
electrochemical reactions can occur that may lead to the
destruction of the sensor.
Preparatory activities
1
Pull the sensor out of the sample and remove any coarse
contamination from the sensor (e.g. brush it off in a bucket of
tap water, wash it down with a hose or wash it off with a cloth).
2
Unscrew the sensor from the SACIQ (SW) sensor connection
cable (see section 5.1 G
).
3
Unscrew the protective hood from the sensor.