2 setting the cell constant manually, Setting the cell constant manually, Lc l e – Xylem handylab LF 11 User Manual
Page 23
handylab LF 11
Operation
23
ba75365e04
01/2011
4.3.2
Setting the cell constant manually
Note
The cell constant to be set must either be taken from the operating manual of
the measuring cell or is printed on the measuring cell.
You can set the cell constant manually as follows:
Range
0.250 ... 2.500 cm
-1
1
Press the
display.
2
Press the
3
Press the
repeatedly until the adjustable cell constant,
e.g. 0.278 cm
-1
appears.
4
Set the cell constant to be used with <V> <W>
, e.g.
0.614 cm
-1
.
O
UpH
L F
Sal
Store
RCL
AR
Auto
AutoCal DIN
AutoCal TEC
LoBat
Cal
ARng
Tref25 Tref20
Lin
nLF
Sal
%
mg/l
mV/pH
cm
°
°
TP
m
S/
cm
cm
1/
F
C
%
/
K
M
TDS
8
1
S
mbar
Time
Baud
Day.Month No.
Ident
Year
L
C L
E
Cal
1
LoBat
Lin
Tref20
RCL
AR
°
°
TP
F
C
%
/
K
M
%
mg/l
mV/pH
m
Sal
TDS
S
Auto
Store
Sal
mbar
O
UpH
cm
AutoCal DIN
AutoCal TEC
Time
Baud
Day.Month No.
Ident
Year
8
0 7
2
2 8
4
cm
1/
S/
cm
Tref25
nLF
ARng
Cal
1
LoBat
Lin
Tref20
RCL
AR
°
°
TP
F
C
%
/
K
M
%
mg/l
mV/pH
m
Sal
TDS
S
Auto
Store
Sal
mbar
O
UpH
cm
AutoCal DIN
AutoCal TEC
Time
Baud
Day.Month No.
Ident
Year
4
0 1
6
8 0
6
cm
1/
S/
cm
Tref25
nLF
ARng