A l c, Lc l e – Xylem handylab LF 11 User Manual
Page 21

handylab LF 11
Operation
21
ba75365e04
01/2011
4.3.1
Determining the cell constant (calibrating in the
control standard)
Note
This method of automatically determining the cell constant by calibration with
the 0.01 mol/l KCL standard solution can only be used for measuring cells
with cell constants in the range 0.450 ... 0.500 cm
-1
or 0.800 ... 1.200 cm
-1
.
This is how you can determine the cell constant:
1
Press the
2
Press the
3
Press the
stant appears on the display: CAL appears on the display.
4
The displayed value is the current, calibrated cell constant. You
can:
– accept this setting for measuring with
– continue with step 5 and start a new calibration.
5
Immerse the measuring cell in the control standard solution,
0.01 mol/KCl.