beautypg.com

3 salinity, 4 tds (total dissolved solids), Salinity – Xylem handylab LF 11 User Manual

Page 18: Tds (total dissolved solids)

background image

Operation

handylab LF 11

18

ba75365e04

01/2011

4.2.3

Salinity

You can measure the salinity as follows:

4.2.4

TDS (Total dissolved solids)

You can measure the total dissolved solids as follows:

1

Perform the preparatory activities according to section 4.2.1.

2

Immerse the conductivity measuring cell in the test sample.

3

Press the key until Sal appears on the status display. The sa-
linity value appears on the display.

4

Wait for a stable measured value.

Tref25

/

K

cm

1/

°

F

%

8

mg/l

1

RCL

ARng

AR

LoBat

Cal

Tref20

Lin

m

S/

cm

M

TDS

S

Auto

Store

Sal

mbar

%

O

UpH

mV/pH

cm

AutoCal DIN

AutoCal TEC

Time

Baud

Day.Month No.

Ident

Year

2

1

2

2 3

3

°

TP

Sal

C

nLF

1

Perform the preparatory activities according to section 4.2.1.

2

Immerse the conductivity measuring cell in the test sample.

3

Press the key until TDS appears on the status display.
The value of the total dissolved solids appears in the upper display
line. The TDS factor appears in the lower display line.

4

Using <V> <W>, set the TDS factor (0.40 ... 1.00).
(The TDS factor has to be determined by a comparison measure-
ment before.)