3 determining/setting up the cell constant [c, Determining/setting up the cell constant [c, Bed in section 4.3 – Xylem handylab LF 11 User Manual

Page 20

background image

Operation

handylab LF 11

20

ba75365e04

01/2011

4.3

Determining/setting up the cell constant [C]

Why determine/set

up the cell

constant?

Aging slightly changes the characteristics of the cell, e. g. by coatings. As a
result, an inexact measured value is displayed. The original characteristics of
the cell can often be restored by cleaning the cell. Calibration determines the
current cell constant and stores it in the instrument.
Thus, you should calibrate at regular intervals (we recommend: every 6
months).

Procedure

The cell constant is determined in the control standard, 0.01 mol/l KCl.

You can determine the actual cell constant of the conductivity measuring cell
by calibrating with the control standard in the following ranges:

z 0.450 ... 0.500 cm

-1

z 0.800 ... 1.200 cm

-1

Besides, you can set the cell constant manually in the following ranges:

z 0.090 ... 0.110 cm

-1

z 0.250 ... 2.500 cm

-1

The fixed cell constant, 0.010 cm

-1

can also be selected. It is not necessary

to calibrate or adjust it.

Cell constants outside the above mentioned ranges cannot be calibrated.

Calibration

evaluation

After the calibration, the measuring instrument automatically evaluates the
current status of the calibration. The evaluation appears on the display.

Display

Cell constant [cm

-1

]

0.450 ... 0.500 cm

-1

0.800 ... 1.200 cm

-1

E3
Eliminate the error according to
chapter 6 W

HAT

TO

DO

IF

...

outside the ranges
0.450 ... 0.500 cm

-1

or
0.800 ... 1.200 cm

-1