Adjustment of 20 ma analog output, 12 ppm – Analytical Industries GPR-7500 AIS Trace PPM Hydrogen Sulfide Analyzer User Manual

Page 27

background image

Advanced Instruments, Inc.

27

4. Advance the reverse shade cursor using the ARROW keys to highlight DEFAULT SPAN.

5. Press the ENTER key to select the highlighted menu option.

The following displays appear and after 3 seconds the system returns to the SAMPLING mode:


Adjustment of 20 mA Analog Output

In rare instances the 4-20mA signal output may not agree to the reading displayed by the LCD. This feature
enables the user to adjust the 20mA signal output should the LCD display not agree.

Note: Adjust the 4mA signal output with the OUTPUT ZERO option described above.

1. Access the MAIN MENU by pressing the MENU key.

2. Advance the reverse shade cursor using the ARROW keys to highlight CALIBRATION.

3. Press the ENTER key to select the highlighted menu option.

The following displays appear:

4. Advance the reverse shade cursor using the ARROW keys to highlight DEFAULT SPAN.

5. Press the ENTER key to select the highlighted menu option.

The following display appears

MAIN MENU

AUTO SAMPLE

MANUAL SAMPLE

CALIBRATION

CONFIG ALARMS

BYPASS ALARMS

CALIBRATION

SPAN CALIBRATE

ZERO CALIBRATE

DEFAULT SPAN

DEFAULT ZERO

OUTPUT SPAN

OUTPUT ZERO

12 PPM

Auto Sampling

0-500 PPM Range

76 F 97 Kpa

LO1 0 PPM HI 40 PPM

FACTORY

DEFAULTS

SET

MAIN MENU

AUTO SAMPLE

MANUAL SAMPLE

CALIBRATION

CONFIG ALARMS

BYPASS ALARMS

CALIBRATION

SPAN CALIBRATE

ZERO CALIBRATE

DEFAULT SPAN

DEFAULT ZERO

OUTPUT SPAN

OUTPUT ZERO