Establishing power to electronics, Manu navigation, Main menu – Analytical Industries GPR-7500 AIS Trace PPM Hydrogen Sulfide Analyzer User Manual
Page 19: 12 ppm

Advanced Instruments, Inc.
19
Establishing Power to Electronics
Once the two power input wires of the shielded cable are properly connected to the terminals inside the Ex
enclosure as described above, connect the other end of the two wires to a suitable 12-24 VDC power source
such as a battery, PLC, DCS, etc.
The digital display responds instantaneously. The transmitter performs several self-diagnostic system status
checks termed as
“START-UP TEST” as illustrated below:
After self diagnostic tests, the analyzer turns itself into the sampling mode. And displays H2S contents the
sensor is exposed to, the analysis range, the ambient temperature and pressure, High and Low alarm set
points.
Manu Navigation
The four (4) pushbuttons located on the front of the transmitter control the micro-processor functions:
Blue ENTER (select)
Yellow UP ARROW
Yellow DOWN ARROW
Green MENU (escape)
Main Menu
To access the MAIN MENU, press the MENU (ESC) key and the following screen will appear.
START-UP TEST
ELECTRONICS
– PASS
TEMP SENSOR
– PASS
BAROMETRIC SENSOR
– PASS
REV 2.40
12 PPM
Auto Sampling
0-500 PPM Range
76 F 97 Kpa
LO1 0 PPM HI 40 PPM
MAIN MENU
AUTO SAMPLE
MANUAL SAMPLE
CALIBRATION
CONFIG ALARMS
ALARM BYPASS
BYPASS ALARMS