beautypg.com

Set-up for manual operation, Set-up for remote operation, Operation – Ocean Optics Multimode Spectrum User Manual

Page 13

background image

2: Set-up and Operation

Set-up for Manual Operation

Procedure

1. Mount laser module onto a suitable heat sink using corner mounting holes.

2. Ensure that external laser enable switch on sidewall of laser module is in the Off (O) position.

3. Plug 5 V DC power supply into wall outlet and then into 5 V DC power input jack on back of

module.

Set-up for Remote Operation

Procedure

1. Mount laser module onto a suitable heat sink using corner mounting holes.

2. Ensure that external laser enable switch on sidewall of laser module is in the Off (O) position.

3. Plug the prewired 10 pin Molex connector into the laser module.

4. Ensure that external 5 V DC power to the module is off (pin 6) and the Laser enable (Pin 7) is

tied to ground.

Note

To use the remote modulation option on this module, connect Pin 2 to Ground and bias
LD SET on Pin 8 between 0 and 1 V to adjust laser output power from 0 to max.

Operation

Read the following warning before attempting to use the laser:

WARNINGS

DO NOT LOOK DIRECTLY INTO THE LASER. LASER RADIATION IS
HAZARDOUS TO THE EYES.

AVOID DIRECT EXPOSURE TO THE BEAM.

000-10000-080-02-0505

5