RSS-VA Installation and Operation Instructions

2

351-A3510-000-01-201010

How the RSS-VA Works

The RSS-VA features a unique internal design that enables it to obtain sample measurements at
varying Angles Of Incidence (AOI). The RSS-VA works as follows:

Light enters the RSS-VA via the SMA 905 input connector (1).

1. Light is reflected up onto the sample port (2) at the degree determined by the AOI setting on

the RSS-VA, where it interacts with the sample.

2. Sampled light is reflected back down through the RSS-VA.

3. Light exits the RSS-VA via the SMA 905 exit connector (3).

Assembling the Hardware

You must first configure your sampling hardware and the RSS-VA prior to measuring your sample.

Procedure

Follow the steps below to set up the RSS-VA:

1. Connect a fiber from the light source to the SMA 905 input connector of the RSS-VA.

2. Connect a fiber from the SMA 905 output connector of the RSS-VA to the spectrometer.

3. Place a STAN-SSH standard on the sample port.

4. Start your spectrometer operating software. Once data acquisition begins, a typical bell-

shaped curve should be displayed in the graph area.

5. Adjust the position of the slide carrier on the RSS-VA until the pointer on the RSS-VA is at

the desired AOI.

6. Adjust the integration time in the software until the signal peaks at approximately 3500

counts.

You have now properly configured the RSS-VA hardware.