RaySphere and RaySphere 1700 Instructions

RAYSPHERE-1700-01-201304

7

Validating the Test Results

Check the following to be sure that you are collecting valid data:

The RaySphere was allowed to warm up for at least 20 minutes before starting the test.

No error or warning messages were received when the calibration file was loaded.

The spectra recording was during the stable flash time phase. Check that all marks for
Start and End of Recording in the Photocurrent graph are at a stable flash time.

Specifications

Specification

Value

Electrical:

Input voltages

90 – 164 VAC

Input frequency

47 – 63 Hz

Power

75 W

Overvoltage protection

110 – 130% of main output

Fuse

M1.6 A

Trigger connection

BNC connector

Connector:

SMA 905

Optical:

Optical input power

150 W to 2000 W

Wavelength range*:

RaySphere

RaySphere 1700

350 – 1100 nm

300 nm – 1700 nm

Optical Trigger:

Trigger delay

0 – 160 ms

Time resolution

0.87 µs

Jitter

150 ns

Internal time resolution

2.54 µs

Electrical Trigger:

Electrical trigger type

High level

High level

5 V or 24 V (selected at the power supply. Galvanically isolated input.)

Minimum high time

100 ns

Connector

BNC