Hg-1 specifications, Calibration procedure, Spectral output – Ocean Optics HG-1 Mercury Argon User Manual
Page 4
![background image](/manuals/292551/4/background.png)
HG-1-CAL Installation and Operation Instructions
4
008-00000-STD-01-20100412
HG-1 Mercury Argon Calibration Light Source
Ocean Optics spectrometer and its manual
An optical fiber (for spectrometers without a built-in slit, a 50-
m fiber works best)
A spreadsheet program (Excel or Quattro Pro, for example) or a calculator that performs third-
order linear regressions
Note
If you are using Microsoft Excel, choose Tools | Add-Ins and check
AnalysisToolPak and AnalysisTookPak-VBA.
Calibration Procedure
Please see the appropriate manual for your model Ocean Optics spectrometer for instructions
on how to calibrate it. Ocean Optics manuals are located at
http://www.oceanoptics.com/technical/operatinginstructions.asp
.
HG-1 Specifications
The following sections detail the specifications of the HG-1.
Spectral Output
Mercury emission lines are <600 nm. Argon emission lines are >600 nm, and are shown
here on an exaggerated amplitude scale.
- Apex 785 Raman (1 page)
- STS-UV (2 pages)
- TR2 Engineering Note (4 pages)
- SAD500 Communications and Control (19 pages)
- Red Tide USB650 Install (26 pages)
- Fiber Optic Termination Kit (6 pages)
- Transmissive pH Probe (10 pages)
- Remora (42 pages)
- PlasCalc (59 pages)
- Correcting Device Driver Issues (8 pages)
- ecoVis Krypton Light Source (16 pages)
- LPC-500CM (28 pages)
- HPX-2000 (24 pages)
- ADC1000-USB (27 pages)
- Torus Operating Instructions (30 pages)
- ADC2000-PCI (13 pages)
- Sensors for Real-Time Analysis (2 pages)
- IDRaman reader (2 pages)
- DH-2000-CAL (30 pages)
- QE65 Pro (32 pages)
- Collimating Lenses (2 pages)
- D1000 (2 pages)
- IDRaman mini (2 pages)
- HR2000CG-UV-NIR (42 pages)
- Cool Red (1 page)
- HL-2000 (20 pages)
- XE-1 Xenon (6 pages)
- USB-ADP Serial Adapters (3 pages)
- CHEM2000 (36 pages)
- Deuterium-Halogen Calibration Light Source (19 pages)
- NeoFox Engineering Note (30 pages)
- QE Pro (2 pages)
- OOIBase32 (140 pages)
- AR-1 Argon (6 pages)
- pH Sensor Patches, Probes and Cuvettes (36 pages)
- SpecLine Offline Spectroscopy (60 pages)
- HL-2000-HP-232 (26 pages)
- External Triggering Options Instructions for Spectrometers with Firmware Version 3.0 and Above (16 pages)
- Breakout Box (10 pages)
- USB-ISS-UV_VIS (4 pages)
- ISS-UV_VIS (6 pages)
- OOIColor (14 pages)
- LS-1 Series (12 pages)
- Apex Install (24 pages)
- DH-2000 (34 pages)