HG-1-CAL Installation and Operation Instructions

4

008-00000-STD-01-20100412

 HG-1 Mercury Argon Calibration Light Source
 Ocean Optics spectrometer and its manual
 An optical fiber (for spectrometers without a built-in slit, a 50-

m fiber works best)

 A spreadsheet program (Excel or Quattro Pro, for example) or a calculator that performs third-

order linear regressions

Note

If you are using Microsoft Excel, choose Tools | Add-Ins and check
AnalysisToolPak and AnalysisTookPak-VBA.

Calibration Procedure

Please see the appropriate manual for your model Ocean Optics spectrometer for instructions
on how to calibrate it. Ocean Optics manuals are located at

http://www.oceanoptics.com/technical/operatinginstructions.asp

.

HG-1 Specifications

The following sections detail the specifications of the HG-1.

Spectral Output

Mercury emission lines are <600 nm. Argon emission lines are >600 nm, and are shown
here on an exaggerated amplitude scale.