HG-1-CAL Installation and Operation Instructions

008-00000-STD-01-20100412

3

Alternately, you can use a 9-volt battery (not included) to power the HG-1. Open the battery
hatch of the HG-1 and install the 9-volt battery, and then proceed to Step 2.

2. Connect a fiber to the SMA 905 Connector on the HG-1. If your spectrometer does not have

an entrance slit, use a 50

m diameter (or smaller) fiber. Larger fibers and slits result in

reduced optical resolution.

Note

If the spectrometer does not have a slit and your experiment requires you to use fibers
of varying diameters, you will need to perform a wavelength calibration after
changing fibers. You should perform a wavelength calibration each time you unscrew
the fiber from the spectrometer and change fiber size.

3. Move the On/Off switch on the HG-1 (next to the SMA 905 Connector) to the On position.

The red LED will illuminate to indicate that the HG-1 is powered on.

You have now configured the HG-1 for use.

Calibrating With the Light Source

The information in this section explains how to calibrate your spectrometer’s wavelength using the
HG-1 light source.

About the Wavelength Calibration

You are going to be solving the following equation, which shows that the relationship between pixel
number and wavelength is a third-order polynomial.

p

= I + C

1

p + C

2

p

2

+ C

3

p

3

Where

= the wavelength of pixel p,

I = the wavelength of pixel 0,
C

1

= the first coefficient (nm/pixel),

C

2

= the second coefficient (nm/pixel

2

)

C

3

= the third coefficient (nm/pixel

3

)

You will be calculating the value for I and the three Cs.

Calibration Requirements for Ocean Optics
Spectrometers

To re-calibrate the wavelength of your Ocean Optics spectrometer using the HG-1, you will need the
following items: