beautypg.com

3B Scientific Magnetic Field Sensor ±2000 mT User Manual

Page 7

background image

3B Scientific GmbH • Rudorffweg 8 • 21031 Hamburg • Germany •

www.3bscientific.com

Subject to technical amendments

© Copyright 2011 3B Scientific GmbH

2 Coils D with 600 Taps

1000988
U8497430

1 Pair of Pole Shoes

1009935
U8497205

1 Set of 15 Safety Experiment Leads

1002843
U138021

Set up the experiment as in Fig. 1.

Switch the transformer and rectifier to the 12-V
setting to supply the basic Hall-effect appara-
tus with 12 V AC.

Select the 2-T range on the magnetic field
sensor and press the Tare button.

Insert the magnetic field sensor into the posi-
tioning hole marked "MFS".

Set the DC power supply to constant-current
mode, i.e. turn the current-setting potentiome-
ter all the way to the left (0 A) and the voltage-

setting potentiometer all the way to the right
(20 V).

Increase the current in the series circuit includ-
ing the transformer coils from 0 A to 2 A in
steps of 0.1 A.

You may also measure the current in the coils
using the current measurement input of the 3B
NETlog

TM

unit (measuring range 2 A DC) and in-

clude those measurements in your evaluation
of the experiment.

Example results:

For a pole-piece separation of 8 mm and values of
10.6 V and 1.74 A the flux density will be 300 mT.


Fig. 1 Measurement of flux density in the air gap between pole pieces of an electro-magnet in an experiment set-up using

the basic Hall-effect apparatus