MT-3109 AC/DC Digital Clamp Multimeter

F. Frequency

Range Resolution Accuracy

Overload

protection

10Hz 0.00Hz

100Hz 0.01Hz

1kHz 0.1Hz

10kHz 1Hz

100kHz 10Hz

1MHz 100Hz

+(0.1 %+5)

10MHz 1kHz reference

only

600Vp

Remark:
Input Sensitivity as follows:
When ≤ 100kHz: ≥ 300mV rms;
When > 100kHz: ≥ 600mV rms
When > 1MHz: ≥ 800mV rms

G. Duty Cycle

Range Resolution Accuracy

Overload

protection

0.1 %

~99.9%

0.1 %

reference

only

600Vp

H. DC Current

Range Resolution Accuracy Overload

protection

40.00A 0.01A +(2.5%+5)
400.0A 0.1A +(2.5%+3)

400A DC/AC

Warning

The operating temperature must be 0 ~40 when measuring

current.
Remark:

 If the reading is positive, the current direction is from bottom to

up. See figure 10, the front case face up while the bottom case
face down. Hold the Meter tight, do now release. The hall
components are very sensitive not only to the magnet but also
to heat and machines reaction force. Any shock will cause
the changing in reading in the short time. Follow the below
procedure to measure current will be more precise: