• Overload Protection: 250V DC or AC Peak Value.

WARNING: DO NOT input any voltage at this range for

safety!


2-2-7. Capacity (F)
A) Rotate function switch to “

“setting.

B) Or rotate the Range to ““setting. Press “SELECT” button to

select the capacity measurement mode.

C) Insert the black test lead to “COM” terminal and red test lead

to “mA“ terminal;

D) Connect the tested capacity by the test leads to “COM”, “mA

“ input terminal, the screen will show capacitance parameter.
(The relative value measurement could be conducted by
pressing “REL” button.)

Caution:

1. Fully discharge the tested capacitor in case it damages the

meter.

2. When measuring in-line capacitor, the power should be turned

off and all capacitors should be discharged completely.

3. It takes about 30 seconds to input stable reading at 200uF

Range.

Range Accuracy Resolution
40nF ±(5.0%+30d)

10pF

400nF ±(3.5%+8d)

100pF

4μF ±(3.5%+8d)

1nF

20μF ±(3.5%+8d)

10nF

WARNING: DO NOT input any voltage at this range for

safety!