Prokit's Industries MT-1217 User Manual
Page 11
2-2-5. Resistance (Ω)
A) Insert the black test lead into “COM” terminal and red test
lead into “
“terminal.
B) Rotate the Range to “Ω” setting. the initial state of the meter
is in automatic range status, which shows “AUTO” symbol
C) Connect the two test leads to the tested resistor.
D) When measuring the low resistance, please short-circuit the
test leads at first to test the wire resistance, and then deduct
it from the actual resistance.
Caution:
1. If “OL” is displayed on LCD, it indicates the tested resistance
value has exceeded the present range limit, please select
higher range to complete the measurement. When measuring
the Resistor higher than 1MΩ, the instrument will take several
seconds to make the reading stable. It is normal when
measuring the high resistor.
2. When the input terminal is open circuit, it will display “OL”.
3. When measuring in-line resistor, be sure that the power is off
and all capacitors are discharged completely.
Range Accuracy Resolution
400Ω ±(0.8%+5d)
0.1Ω
4kΩ
1Ω
40kΩ 10Ω
400kΩ 100Ω
4MΩ
±(0.8%+4d)
1kΩ
40MΩ ±(1.2%+10d)
10kΩ
• Open Voltage circuit: Less than 200mV;
• Overload Protection: 250V DC or AC Peak Value;
Note: When measuring at Range 400Ω, please short-circuit the
test leads at first to test the wire resistance, and then deduct
it from the actual resistance.
10