VEGA VEGAPULS 61 (≥ 2.0.0 - ≥ 4.0.0) Profibus PA Approval according to LPR radio standard User Manual
Page 69

0,5 m (1.6 ft)
10 mm (0.394 in)
- 10 mm (- 0.394 in)
2 mm (0.079 in)
- 2 mm (- 0.079 in)
0
1
2
3
Fig. 51: Deviation under reference conditions - plastic horn antenna
1
Reference plane
2
Antenna edge
3
Recommended measuring range
Reproducibility
≤ ±
1 mm
Deviation with bulk solids
The values depend to a great extent on the
application. Binding specifications are thus not
possible.
Variables influencing measurement accuracy
Temperature drift - Digital output
±
3 mm/10 K relating to the max. measuring range
or max. 10 mm
Additional deviation through strong, high
frequency electromagnetic fields acc. to EN
61326
< ±
50 mm
Characteristics and performance data
Frequency
K-band (26 GHz technology)
Measuring cycle time
450 ms
Step response time
1)
≤
3 s
Tracking speed of the measuring window
max.
1 m/min
Beam angle
2)
10°
Emitted HF power
3)
-
Average spectral transmission power
density
-34 dBm/MHz EIRP
1)
Time span after a sudden measuring distance change by max. 0.5 m in
liquid applications, max 2 m with bulk solids applications, until the output
signal has taken for the first time 90 % of the final value (IEC 61298-2).
2)
Outside the specified beam angle, the energy of the radar signal has a level
which is reduced by 50 % (-3 dB)
3)
EIRP: Equivalent Isotropic Radiated Power
VEGAPULS
61 • Approval according to LPR radio standard
69
11 Supplement
41715
-
EN
-120301