4 Mounting

VEGASWING 66 • - two-wire

44952-EN-130805

The max. permissible pressure is specified in chapter "Technical

data" or on the type label of the sensor.
Note:

Seal for instruments with process fitting thread
The thread and the seal form on the mounting boss correspond to

DIN 3852, part 1, screwed plug Form B (sealing via metallic sealing

edge). In this case, no seal is required.

Instruments for high temperatures have a temperature adapter be-

tween process fitting and electronics housing. This is used for thermal

decoupling of the electronics from high process temperatures.
Information:

The temperature adapter may be embedded in the vessel insulation

only up to max. 50 mm (1.97 in). Only then is a reliable temperature

decoupling guaranteed.

1

2

Fig. 8: Mounting the instrument on insulated vessels.
1 Temperature isolation - max. 50 mm (1.97 in)

2 Ambient temperature on the housing

If VEGASWING 66 is mounted in the filling stream, unwanted false

measurement signals can be generated. For this reason, mount VE-

GASWING 66 at a position in the vessel where no disturbances, e.g.

from filling openings, agitators, etc., can occur.
This applies particularly to instrument types with long extension tube.

Mounting in the vessel

insulation

Inflowing medium