6 Set up with the display and adjustment module
VEGABAR 83 • 4 … 20 mA/HART
45034-EN-131010
If you have connected no slave sensor, you confirm this with "Deac-
tivate".
VEGABAR 83 can be used for process pressure and level measure-
ment. Default setting is process pressure measurement. The mode
can be changed in this adjustment menu.
The VEGABAR 83 in conjunction with a slave sensor can be used for
flow, differential pressure, density and interface measurement. The
default setting is differential pressure measurement. Switchover is
carried out in the adjustment menu.
Depending on the selected application, different subchapters in the
following adjustment steps are important. There you can find the
individual adjustment steps.
Enter the requested parameters via the appropriate keys, save your
settings with [OK] and jump to the next menu item with the [ESC] and
the [->] key.
In this menu item, the adjustment units of the instrument are deter-
mined. The selection determines the displayed unit in the menu items
"Min. adjustment (zero)" and "Max. adjustment (span)".
Unit of measurement:
If the level should be adjusted in a height unit, the density of the me-
dium must also be entered later during the adjustment.
In addition, the temperature unit of the instrument is specified. The se-
lection determines the displayed unit in the menu items "Peak value,
measuring cell temperature", "Peak value, electronics temperature"
as well as "HART variables".
Temperature unit:
Enter the requested parameters via the appropriate keys, save your
settings with [OK] and jump to the next menu item with the [ESC] and
the [->] key.
Especially with chemical seal systems, the installation position of the
instrument can shift (offset) the measured value. The position cor-
rection compensates this offset. Hence the actual measured value is
taken over automatically. With relative pressure measuring cells also a
manual offset can be carried out.
Setup - Units
Setup - Position correc-
tion