4 Mounting

VEGABAR 83 • 4 … 20 mA/HART

45034-EN-131010

1

Fig. 9: Position of the filter element - IP 69K version
1 Filter element

Instruments with absolute pressure have a blind plug mounted

instead of the filter element.

4.3 Process pressure measurement

Keep the following in mind when setting up the measuring system:

Mount the instrument above the measuring point

Possible condensation can then drain off into the process line.

1

2

3

Fig. 10: Measurement setup for process pressure measurement of gases in

pipelines
1 VEGABAR 83

2 Blocking valve

3 Pipeline

Keep the following in mind when setting up the measuring system:

Connect via a siphon

Do not insulate the siphon

Fill the siphon with water before setup

Measurement setup in

gases

Measurement setup in

vapours