VEGA VEGAFLEX 83 Modbus and Levelmaster protocol Polished version User Manual
Page 83
83
11 Supplement
VEGAFLEX 83 • Modbus and Levelmaster protocol
41843-EN-130314
•
All commands start with "Uuu", whereby "uu" stands for the address (00-31)
•
"*" can be used as a joker for any position in the address. The sensor always converts this in
its address. In case of more than one sensor, the joker must not be used, because otherwise
several slaves will answer
•
Commands that modify the instrument return the command with "OK“. "EE-ERROR" replaces
"OK" if there was a problem changing the configuration
Report Level (and Temperature)
Request:
Parameter
Length
Code/Data
Report Level (and Temperature)
4 characters ASCII
Uuu?
Response:
Parameter
Length
Code/Data
Report Level (and Temperature)
24 characters ASCII
UuuDlll.llFtttEeeeeWwwww
uu = Address
lll.ll = PV in inches
ttt = Temperature in Fahrenheit
eeee = Error number (0 no error, 1
level data not readable)
wwww = Warning number (0 no
warning)
Level in inches is repeated for sensors with 2 levels. PV value is transmitted as first measured value,
SV as second measured value.
PV, SV, TV and QV can be adjusted via the sensor DTM.
Report Unit Number
Request:
Parameter
Length
Code/Data
Report Unit Number
5 characters ASCII
U**N?
Response:
Parameter
Length
Code/Data
Report Level (and Temperature)
6 characters ASCII
UuuNnn
Assign Unit Number
Request:
Parameter
Length
Code/Data
Assign Unit Number
6 characters ASCII
UuuNnn
Response:
Parameter
Length
Code/Data
Assign Unit Number
6 characters ASCII
UuuNOK
uu = new Address