VEGA VEGAFLEX 83 Modbus and Levelmaster protocol Polished version User Manual
Page 11
11
3 Product description
VEGAFLEX 83 • Modbus and Levelmaster protocol
41843-EN-130314
2
1
Fig. 5: Position of Modbus and sensor electronics
1 Modbus electronics
2 Sensor electronics
This operating instructions manual applies to the following instrument
versions:
•
Hardware from 1.0.0
•
Software from 1.0.0
•
Only for instrument versions without SIL qualification
The respective electronics version can be determined via the product
code on the type label as well as on the electronics.
•
Standard electronics: Type FX80HK.-
The scope of delivery encompasses:
•
Sensor
•
Documentation
– this operating instructions manual
– Test certificate measuring accuracy (optional)
– Operating instructions manual "Display and adjustment mod-
ule" (optional)
– Supplementary instructions "GSM/GPRS radio module"
(optional)
– Supplementary instructions manual "Heating for display and
adjustment module" (optional)
– Supplementary instructions manual "Plug connector for con-
tinuously measuring sensors" (optional)
– Ex-specific "Safety instructions" (with Ex versions)
– if necessary, further certificates
3.2 Principle of operation
The VEGAFLEX 83 is a level sensor with polished rod probe for
continuous level or interface measurement, particularly suitable for
applications in the food processing and pharmaceutical industry.
Optionally an autoclaved version with separable housing is available.
High frequency microwave pulses are guided along a steel cable or
a rod. Upon reaching the product surface, the microwave pulses are
reflected. The running time is evaluated by the instrument and output-
ted as level.
Scope of this operating
instructions manual
Versions
Scope of delivery
Application area
Functional principle -
level measurement