2 communication profibus pa – VEGA VEGAFLEX 81 Profibus PA User Manual
Page 90

90
11 Supplement
VEGAFLEX 81 • Profibus PA
44217-EN-130910
Ʋ Inspection window
Polyester foil
Integrated clock
Date format
Day.Month.Year
Time format
12 h/24 h
Time zone Ex factory
CET
Rate deviation max.
10.5 min/year
Measurement electronics temerature
Resolution
1 °C (1.8 °F)
Accuracy
±1 °C (1.8 °F)
Voltage supply
Operating voltage
Ʋ Non-Ex instrument
9 … 32 V DC
Ʋ EEx-ia instrument - Power supply
FISCO model
9 … 17.5 V DC
Ʋ EEx-ia instrument - Power supply
ENTITY model
9 … 24 V DC
Ʋ EEx-d instrument
14 … 32 V DC
Number of sensors per DP/PA segment coupler, max.
Ʋ Non-Ex
32
Ʋ Ex
10
Electrical protective measures
Protection, depending on housing version
Ʋ Plastic housing
IP 66/IP 67
Ʋ Aluminium housing, stainless steel
housing - investment casting, stain-
less steel housing - electro-polished
IP 66/IP 68 (0.2 bar)
5)
Ʋ Aluminium and stainless housing, in-
vestment casting (optionally available)
IP 66/IP 68 (1 bar)
Overvoltage category
III
Protection class
III
Approvals
Instruments with approvals can have different technical data depending on the version.
For that reason the associated approval documents of these instruments must be carefully noted.
They are part of the delivery or can be do
w
11.2 Communication Profibus PA
Instrument master file
The instrument master file (GSD) contains the characteristic data of the Profibus PA instrument.
5)
The prerequisites for maintaining the protection rating are a suitable cable as well as correct mounting.