6 Set up with the display and adjustment module

SOLITRAC 31 • Profibus PA

43834-EN-131119

The address setting is carried out either via:

The address selection switch in the electronics compartment of

the instrument (address setting via hardware)

The display and adjustment module (address setting via software)

PACTware/DTM (address setting via software)

The hardware addressing is effective if an address <126 is set

with the address selection switches on the electronics module of

SOLITRAC 31. Software addressing thus has no effect - only the set

hardware address applies.

The software addressing is only effective if address 126 or higher is

adjusted on the instrument with the address selection switches.

If a hardware address was adjusted (< 126), then the instrument

signals that the hardware address is active.

Here you can assign a suitable measurement loop name. Push the

"OK" key to start the processing. With the "+" key you change the sign

and with the "->" key you jump to the next position.
You can enter names with max. 19 characters. The character set

comprises:

Capital letters from A … Z

Numbers from 0 … 9

Special characters + - / _ blanks

In this menu item you can adjust the SOLITRAC 31 to the integrated

isotope in the source container.
For this purpose, check which isotope is integrated in the source

container. You can find this information on the type label of the source

container.

With this selection, the sensitivity of the sensor is adapted in an opti-

mum way to the isotope.

Hardware addressing

Software addressing

Setup - Measurement

loop name

Setup - Isotope