beautypg.com

10 diagnostics and service, 1 maintenance, 2 diagnosis memory – VEGA VEGAFLEX 81 4 … 20 mA_HART two-wire Coax probe With SIL qualification User Manual

Page 61

background image

61

10 Diagnostics and service

VEGAFLEX 81 • 4 … 20 mA/HART two-wire

44216-EN-130823

10 Diagnostics and service

10.1 Maintenance

If the device is used correctly, no maintenance is required in normal

operation.
When used in safety-instrumented systems (SIS), the safety function

must be carried out on the instrument in regular time intervals by

means of a proof test.
Hence possible undetected, dangerous failure can be identified.
The operator's responsibility to select the kind of test. The time inter-

vals depend on the used PFD

AVG

.

During the function test, the safety function must be treated as unsafe.

Keep in mind that the function test influences downstream connected

devices.
If one of the tests proves negative, the entire measuring system must

be switched out of service and the process held in a safe state by

means of other measures.
You can find detailed information on the proof test in the Safety

Manual (SIL).

10.2 Diagnosis memory

The instrument has several memories which are available for diagno-

sis purposes. The data remain even with voltage interruption.

Up to 100,000 measured values can be stored in the sensor in a ring

memory. Each entry contains date/time as well as the respective

measured value. Storable values are for example:

Distance

Filling height

Percentage value

Lin. percent

Scaled

Current value

Meas. reliability

Electronics temperature

When the instrument is shipped, the measured value memory is

active and stores distance, measurement certainty and electronics

temperature every 3 minutes.
In "Extended adjustment" you can select the respective measured

values.
The requested values and recording conditions are set via a PC with

PACTware/DTM or the control system with EDD. Data are thus read

out and also reset.

Up to 500 events are automatically stored with a time stamp in the

sensor (non-deletable). Each entry contains date/time, event type,

event description and value. Event types are for example:

Modification of a parameter

Measured value memory

Event memory