7 Set up with the display and adjustment module
VEGAFLEX 86 • 4 … 20 mA/HART two-wire
44234-EN-130823
The respective min. and max. measured values are saved in the
sensor. The values are displayed in the menu item "Peak values Ad-
ditional".
This menu item displays the peak values of the electronics tempera-
ture as well as the dielectric constant.
In another window you can carry out a reset of the two peak values
separately.
The menu item "Echo curve" shows the signal strength of the echoes
over the measuring range in V. The signal strength enables an evalua-
tion of the quality of the measurement.
With the following functions you can zoom part sections of the echo
curve.
•
"X-Zoom": Zoom function for the meas. distance
•
"Y-Zoom": 1, 2, 5 and 10x signal magnification in "V"
•
"Unzoom": Reset the presentation to the nominal measuring range
with single magnification
In this menu item you can simulate measured values via the current
output. This allows the signal path to be tested, e.g. via downstream
indicating instruments or the input card of the control system.
Select the requested simulation size and adjust the requested value.
Diagnostics - Peak values
Additional
Diagnostics - Echo curve
Diagnosis - Simulation