6 Set up the sensor with the indicating and adjustment module

VEGAPULS 65 • Modbus and Levelmaster protocol

41365-EN-121011

Setup: Settings, for example, for medium, application, vessel, adjust-

ment, damping
Display: Language setting, settings for the measured value indication

as well as lighting
Diagnosis: Information, e.g. on the instrument status, pointer, reli-

ability, simulation, echo curve
Further settings: e.g. instrument units, unit SV 2, false signal sup-

pression, linearization, date/time, reset, copy sensor data
Info: Instrument name, hardware and software version, calibration

date, device ID, instrument features
In the main menu point "Setup", the individual submenu points should

be selected subsequently and provided with the correct parameters

to ensure the optimum adjustment of the measurement. The proce-

dure is described in the following.
Setup
Each medium has different reflection properties. With liquids, further

interfering factors are fluctuation product surface and foam genera-

tion. With bulk solids, these are dust generation, material cone and

additional echoes from the vessel wall.
To adapt the sensor to these different measuring conditions, the

selection "Liquid" or "Bulk solid" should be made in this menu item.

Through this selection, the sensor is adapted perfectly to the product

and measurement reliability, particularly in products with poor reflec-

tive properties, is considerably increased.
Enter the requested parameters via the appropriate keys, save your

settings with [OK] and jump to the next menu item with the [ESC] and

the [->] key.

In addition to the medium, also the application, i.e. the measuring site,

can influence the measurement.
With this menu item, the sensor can be adapted to the applications.

The adjustment possibilities depend on the selection "Liquid" or "Bulk

solid" under "Medium".

Setup/Medium