4 Mounting

VEGAPULS 65 • Modbus and Levelmaster protocol

41365-EN-121011

1

Fig. 6: Position of the polarisation level
1 Marking hole

When mounting the VEGAPULS 65, keep a distance of at least

500 mm (19.69 in) to the vessel wall. If the sensor is installed in the

center of dished or round vessel tops, multiple echoes can arise.

These can, however, be suppressed by an appropriate adjustment

(see chapter "Setup").
If you cannot keep this distance you should carry out a false echo

storage before setup. This applies mainly if buildup on the vessel wall

is expected. In this case, we recommend repeating a false echo stor-

age later with existing buildup.

> 500 mm

(19.69")

Fig. 7: Mounting of the radar sensor on round vessel tops

In vessels with conical bottom it can be advantageous to mount the

sensor in the center of the vessel, as measurement is then possible

down to the lowest point of the vessel bottom.

Installation position