beautypg.com

2 principle of operation – VEGA VEGAPULS 65 (≥ 2.0.0 - ≥ 4.0.0) Modbus and Levelmaster protocol User Manual

Page 11

background image

11

3 Product description

VEGAPULS 65 • Modbus and Levelmaster protocol

41365-EN-121011

2

1

Fig. 5: Position of Modbus and sensor electronics
1 Modbus electronics

2 Sensor electronics

This operating instructions manual applies to the following instrument

versions:

Hardware from 2.1.0

Software from 4.5.1

The scope of delivery encompasses:

Radar sensor

Documentation

– this operating instructions manual

– Test certificate Measurement accuracy, depending on the

instrument VEGAPULS 65 (optional)

– Operating instructions manual "Indicating and adjustment mod-

ule PLICSCOM" (optional)

– Supplementary instructions manual "Heating for indicating and

adjustment module" (optional)

– Supplementary instructions manual "Plug connector for con-

tinuously measuring sensors" (optional)

– Ex-specific "Safety instructions" (with Ex versions)

– if necessary, further certificates

3.2 Principle of operation

The VEGAPULS 65 is a radar sensor for continuous measurement of

liquids under simple process conditions. It is particularly suitable for

level measurement in vessels with small process fittings and under

simple process conditions.
The instrument can be used with products with an ε

r

value ≥1.8. The

actually achievable value depends on the measuring conditions, the

antenna system, the standpipe or bypass.

The antenna of the radar sensor emits short radar pulses with a

duration of approx. 1 ns. These pulses are reflected by the product

and received by the antenna as echoes. The transit time of the radar

pulses from emission to reception is proportional to the distance and

hence to the level. The determined level is converted into an appropri-

ate output signal and outputted as measured value.

Scope of this operating

instructions manual

Scope of delivery

Application area

Functional principle