11 Supplement

VEGAPULS 61 • Modbus and Levelmaster protocol

41361-EN-121011

Response:

Parameter

Length

Code/Data

Set Receive to Transmit Delay

5 characters ASCII

UuuFn
n = number of measurement values

(0, 1 or 2)

11.5 Report Receive to Transmit Delay

Request:

Parameter

Length

Code/Data

Report Receive to Transmit Delay

4 characters ASCII

UuuR

Response:

Parameter

Length

Code/Data

Report Receive to Transmit Delay

7 characters ASCII

UuuRmmm
mmm = milliseconds (50 bis 250),

default = 127 ms

11.5 Error codes

Error Code

Name

EE-Error

Error While Storing Data in EEPROM

FR-Error

Erorr in Frame (to short, to long, wrong data)

LV-Error

Value out of limits