3 scale the indication – VEGA VEGAWELL 52 4 … 20 mA_HART Pt 100 User Manual
Page 23
I
f you want to change this value, edit the value via "OK" and adjust the
requested value via [↑] and [↓]. Save the value via "OK", the
VEGAWELL
52
displays briefly "Wait", then the value is written into the
sensor.
M
ove further to the menu item "MB end", there the measuring range
end is displayed, for example the default setting 0.4 bar.
P
roceed in the same way for "MB end, by entering for example the
value 0.2 bar and save as described before.
T
he min./max. adjustment is thereby finished. After "[ESC]" is
pressed, the display shows in the measured value display the currently
measured level in bar (digital) and on the bargraph (quasianalogue).
6
.3 Scale the indication
O
ften the level should not be display in bar but in %.
F
or this purpose, push "OK" to reach the adjustment menu. Select with
[↑]
and [↓] the submenu "Indication" and confirm with "OK". Then
move to the menu item "Unit". There the actual indicating unit of the
sensor and the VEGAWELL 52 is displayed, for example, "bar".
E
dit the indicating unit, select % and save the value with "OK".
A
fter "[ESC]", the display shoes the level in % (digital) on the
measured value indication and on the bargraph (quasianalog).
VEGAWELL
52 • 4 … 20
mA/HART Pt
23