VEGA VEGADIF 65 Profibus PA User Manual
Page 73

73
11 Supplement
VEGADIF 65 • Profibus PA
36129-EN-130417
Max. torque screws mounting strap
30 Nm
Max. torque screws socket external
housing
5 Nm (3.688 lbf ft)
Weight approx.
4.2 … 4.5 kg (9.26 … 9.92 lbs), depending on process
fitting
Output variable
Output signal
digital output signal, format according to IEEE-754
Sensor address
126 (default setting)
Current value
10 mA, ±0.5 mA
Dynamic behaviour output
Run-up time
≤ 20 s
I
63 %
100 %
t
t
1
t
2
Fig. 58: Presentation of the dead time t
1
and the time constant t
2
Version, nominal measuring range
Dead time t
1
Time constant t
2
Basic version 10 bar and 30 bar
100 ms
250 ms
Basic version 100 mbar
100 ms
180 ms
Basic version 500 mbar
100 ms
180 ms
Basic version, 3 bar
100 ms
180 ms
Basic version 16 bar and 40 bar
100 ms
180 ms
Chemical seal version, all nominal measuring ranges
Dependent on the chemi-
cal seal
Dependent on the chemi-
cal seal
Bus response time with Profibus PA
Ʋ cyclically approx.
10 ms
Ʋ acyclically approx.
50 ms
Damping (63 % of the input variable)
0 … 999 s, adjustable
Additional output parameter - temperature
Processing is made via output signal HART multidrop, Profibus PA and Foundation Fieldbus
Range
-50 … +150 °C (-58 … +302 °F)
Resolution
1 °C (1.8 °F)