8 Setup

VEGADIF 65 • Profibus PA

36129-EN-130417

Fig. 53: Preferred measurement setup for open vessels
I VEGADIF 65

II Separator

1 Drain valve

6.7 Vent valves on VEGADIF 65

A Blocking valve

Proceed as follows:
1. Fill the vessel to just over the lower tap.
2. Fill measuring system with medium.

Open valve A: Medium flows in.

3. Vent instrument

Briefly open valve 6, then close it: Fill the measuring instrument

completely with the medium and remove air.

4. Set measurement loop to operation

Now:
Valve A open and valve 6 closed

Then carry out adjustment, see below.

Measurement setup for

open vessels

Prepare the adjustment