8
.3 Remove interferences
The operator of the system is responsible for taking suitable measures
to rectify faults.
Connect a handmultimeter in the suitable measuring range according
to the wiring plan. The following table describes possible errors in the
current signal and helps to remove them:
Error
Cause
Rectification
4 … 20 mA signal
not stable
l Level fluctua-
tions
l Set damping according to the in-
strument via the indicating and
adjustment module or PACTware/
DTM
4 … 20 mA signal
missing
l Electrical con-
nection faulty
l Check connection according to
chapter "Connection steps" and if
necessary, correct according to
chapter "Wiring plan".
l Voltage supply
missing
l Check cables for breaks; repair if
necessary
l Operating vol-
tage too low or
load resis-
tance too high
l Check, adapt if necessary
Current signal
greater than 22 mA
or less than
3.6 mA
l Electronics
module in the
sensor defecti-
ve
l Exchange the instrument or send it
in for repair
Depending on the reason for the fault and the measures taken, the
steps described in chapter "Setup" must be carried out again or must
be checked for plausibility and completeness.
Should these measures not be successful, please call in urgent cases
the VEGA service hotline under the phone no. +49 1805 858550.
The hotline is also available outside the normal working hours on
seven days a week around the clock.
Since we offer this service worldwide, the support is in the English
language. The service itself is free of charge, the only costs involved
are the normal call charges.
8
.4 How to proceed in case of repair
If a repair is necessary, please proceed as follows:
Reaction when malfunc-
tions occur
Check the 4 … 20 mA
signal
Reaction after fault rec-
tification
24
hour service hotline
40
VEGADIS
62
8 Diagnosis and service
36469
-EN
-120329