beautypg.com

VEGA VEGAPULS 61 (≥ 2.0.0 - ≥ 4.0.0) Profibus PA User Manual

Page 23

background image

23

4 Mounting

VEGAPULS 61 • Profibus PA

36501-EN-121011

The tube diameter must be at least DN 40 or 1½" with antenna

size 40 mm (1½")

For the parameter adjustment, select "Application standpipe" and

enter the tube diameter to compensate errors due to running time

shift

A false signal suppression with integrated sensor is recommended

but not mandatory

The measurement through a ball valve with complete run is pos-

sible

Constructional requirements on the surge pipe:

Material metal, smoother inner tube

Preferably pultruded or straight beaded stainless steel tube

Welded joint should be straight and lie in one axis with the holes

Flanges are welded to the tube according to the orientation of the

polarisation level

In case of a extension with a welding neck flange or pipe collar

as well as when a ball valve is used, the inner surfaces should be

aligned and accurately joined together

Gap size with junctions ≤ 0.1 mm

Do not weld through the pipe wall. The surge pipe must remain

smooth inside. Roughness and beads on the inside caused by

unintentional penetration should be removed since they cause

strong false echoes and encourage buildup

Surge pipes must extend all the way down to the requested min.

level, as measurement is only possible within the tube

Diameter of holes ≤ 5 mm, any number OK, on one side or com-

pletely through

The antenna diameter of the sensor should correspond to the

inner diameter of the tube

Diameter should be constant over the complete length

An alternative to measurement in a surge pipe is measurement in a

bypass tube outside of the vessel.

Measurement in the

bypass