VEGA VEGABAR 66 Profibus PA User Manual
Page 55

55
10 Supplement
VEGABAR 66 • Profibus PA
36739-EN-130321
The specifications are only an overview and refer to the measuring cell. Limitations due to the ma-
terial and version of the process fitting are possible. The specifications on the nameplate apply.
Nominal range
Overload capacity, max.
pressure
Overload capacity, min. pres-
sure
Gauge pressure
0 … +1.5 psig
+200 psig
-3 psig
0 … +3 psig
+290 psig
-6 psig
0 … +6 psig
+430 psig
-12 psig
0 … +15 psig
+500 psig
-15 psig
0 … +35 psig
+700 psig
-15 psig
0 … +70 psig
+950 psig
-15 psig
0 … +150 psig
+1300 psig
-15 psig
0 … +350 psig
+1900 psig
-15 psig
0 … +900 psig
+2900 psig
-15 psig
Absolute pressure
0 … 15 psi
500 psi
0 psi
0 … 35 psi
700 psi
0 psi
0 … 70 psi
900 psi
0 psi
0 … 150 psi
1300 psi
0 psi
0 … 350 psi
1900 psi
0 psi
Reference conditions and actuating variables (according to DIN EN 60770-1)
Reference conditions according to DIN EN 61298-1
Ʋ Temperature
+15 … +25 °C (+59 … +77 °F)
Ʋ Relative humidity
45 … 75 %
Ʋ Air pressure
860 … 1060 mbar/86 … 106 kPa (12.5 … 15.4 psig)
Determination of characteristics
Limit point adjustment according to IEC 61298-2
Characterstic curve
Linear
Reference installation position
upright, diaphragm points downward
Influence of the installation position
< 0.2 mbar/20 Pa (0.003 psig)
Deviation determined according to the limit point method according to IEC 60770
11)
Applies to the digital signal output (HART, Profibus PA, Foundation Fieldbus) as well as to ana-
logue current output 4 … 20 mA and refers to the set span. Turn down (TD) is the ratio nominal
measuring range/set span.
Deviation
Ʋ Turn down 1 : 1 up to 5 : 1
< 0.1 %
Ʋ Turn down > 5 : 1
< 0.02 % x TD
Influence of the product or ambient temperature
Thermal change zero signal and output span
11)
Incl. non-linearity, hysteresis and non-repeatability.