beautypg.com

VEGA VEGADIF 55 Service instructions User Manual

Page 9

background image

Code

Message
type/NA
64

Message/De-
scription

Cause

Measure

Priority

604 (W604) Warning C

C> Linearisation
too less points or
points too close

l

The linearization chart
consists of less than
two points. At least
two points of the lin-
earization chart are
too close together. A
min. distance of 0.5 %
of the span between
two points must be
maintained. Spans for
the option "Pressure
with characteristics
":
HYDR. PRSSURE
MAX. – HYDR.
PRESSURE MIN.;
TANK CONTENTS

MAX. – TANK CON-
TENTS MIN. Spans
for option "Height with
characteristics
": LEV-
EL MAX. – LEVEL
MIN.; TANK CON-
TENTS MAX. – TANK
CONTENTS MIN.

l

Complete linearization
chart, if necessary ac-
cept linearization chart
again. Correct lineari-
zation chart and ac-
cept again.

58

613 (W613) Warning I

I> Simulation ac-
tive

l

Simulation switched
on, i.e. the instrument
is not measuring at the
moment

l

Switch off simulation

60

620 (E620)

Error C
factory set-
ting warn-
ing

C> Current out-
side the nominal
range

The current is outside the
permissible range
3.8 … 20.5 mA

l

The existing pressure
is outside the set
measuring range
(however within the
sensor range)

l

Loose contact on the
sensor cable

l

Check the existing
pressure, if necessary
reset the measuring
range (see operating
instructions manual
VEGADIF 55)

l

Carry out reset (Code
7864) and make a
fresh adjustment

l

Wait a short moment
and ensure a stable
connection or avoid
loose connection.

49

700 (W700) Warning C

C> Last configu-
ration not accep-
ted

l

During writing or read-
ing of configuration
data an error occurred
or the voltage supply
was interrupted

l

Main electronics de-
fective

l

Carry out reset (Code
7864) and make a
fresh adjustment

l

Exchange the main
electronics

51

Service instructions VEGADIF 55

9

Maintenance and fault rectification

31732

-

EN

-080430